The Realy Novel Coronavirus (SARS-Cov-2) Antigen Rapid Test Cassette is a rapid chromatographic immunoassay for the qualitative detection of specific antigens in Sars-Cov-2 present in the human nasopharynx.
How to:
Allow test, sample, extraction buffer to come to room temperature
(15-30 ° C) before testing.
1.Remove the tester from the sealed aluminum case and use it
as soon as possible. Place the tester on a clean and flat surface. The
better results will be obtained if the analysis is performed immediately after opening
of the aluminum case.
2. Unscrew the entire cap of the sample collection tube
3. Take out 1 bottle of sample extraction buffer, remove the bottle cap,
add all of the extraction buffer to the extraction tube.
4.Place the sterile swab specimen in the specimen extraction buffer.
Rotate the stylus for about 10 seconds while pressing the head in
of the tube to release the antigen on the swab.
5.Remove the sterile swab while pressing the sterile swab head into
inside the tube as you remove it to expel as much liquid as possible
from the stylus. Dispose of the sterile swab according to the disposal protocol
biohazard waste.
6. Screw and tighten the cap on the sample collection tube, then
vigorously shake the sample collection tube to mix the sample and the
sample extraction buffer. See figure 4.
7.Add 3 drops of solution (about 80μl) to the sample port of the device and to
then start the timer. Read the result in 10~20 minutes. Do not interpret
result after 20 minutes.
Security measures:
For in vitro diagnostic use only.
•Do not use after the expiry date.
• Make sure the test package is not damaged before opening for use.
•Perform the test at room temperature 15 to 30°C.
•Wear gloves when handling samples, avoid touching membrane
reagent and the sample window.
• All samples and used components should be treated as
infectious and to be disposed of according to local regulations.
•Avoid the use of blood samples.
Performance Characteristics:
Clinical evaluation
A clinical evaluation was performed to compare the results obtained from
the Novel Coronavirus (SARS-Cov-2) Rapid Antigen Test Cassette (Pen) and PCR. The
results are summarized below:
Table: : Novel Coronavirus (SARS-Cov-2) Antigen Rapid Test Cassette (swab) vs. PCR
Method
2019 Test K-nCoV it (RT-NPCR ucleic ) Acid
Total results
The Novel Coronavirus (SARS-Cov-2) Antigen Rapid Test Cassette (Swab)
Results
Positive
Negative
Positive
201
0
201
Negative
8
450
458
Overall results
209
450
659
Clinical sensitivity = 201/209=96.17 % (95%CI* 92.51% to 98.17%)
Clinical specificity = 450/450>99.9% (95%CI* 98.98% to 100%)
Accuracy: (201+450)/ (201+0+8+450) *100%=98.79% (95%CI* 97.58% to 99.43%)
*Confidence interval
This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.